Rki-instruments SDM-2012 PC Controlled Configuration Manuel d'utilisateur Page 91

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 218
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 90
Bump Testing a Gas Tracer 86
16. After air is applied, and if the ppm sensor passed the zero adjustment,
calibration gas from the GAS 2 cylinder will be applied to the ppm sensor. The
GAS 2 indicator on the program screen will come on and calibration gas will be
applied to the ppm sensor for the duration specified in the Gas Exposure Time
parameter in the Configuration Window.
Figure 66: GAS 2 PPM Gas Flowing
The PC Controller program will analyze the results. If the bump test on the
ppm sensor fails and Force Calibration is selected, a calibration will
automatically begin on the ppm sensor and GAS 2 calibration gas will continue
to flow.
Vue de la page 90
1 2 ... 86 87 88 89 90 91 92 93 94 95 96 ... 217 218

Commentaires sur ces manuels

Pas de commentaire