Calibrating GX-2012s and Gas Tracers 198
4. Ensure that the gas out to instrument line is connected to the inlet of each
instrument and that the exhaust line is connected to the exhaust fitting of each
instrument.
5. Select the instruments you wish to calibrate and click on them to highlight them
as shown below. To select multiple instruments for calibration, hold down the
Control button on the keyboard as you click them.
Figure 157: Instrument Selection
6. Click the Calibration button in the lower right corner of the computer program
screen.