Rki-instruments GX-2012 Manuel d'utilisateur Page 16

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 90
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 15
12 • Control Buttons
Last Calibration Button
Clicking the Last Calibration button opens the Last Calibration Window.
NOTE: The setting of the BUMP DSP instrument parameter affects
whether the Bump Test portion of the Last Calibration Window is
shown. For further explanation of the BUMP DSP parameter as it
pertains to the Last Calibration Window, see “Effects of BUMP
DSP Parameter Setting on Appearance of Last Calibration
Window” on page 64.
The Last Calibration Window stores the information for the most recent
successful calibration and for the most recent successful bump test for
each instrument that has been downloaded. Calibration data is displayed in
the top part of the screen while bump test data is displayed in the bottom
part. You can display the information three ways by using the Need
Calibration/Need Bump Test, Calibration Date/Bump Test Date, or
Calibration Record/Bump Test Record selection buttons. You can also print
the information if you select the Need Calibration/Need Bump Test or
Calibration Date/Bump Test Date display options.
Figure 7: Last Calibration Window
Vue de la page 15
1 2 ... 11 12 13 14 15 16 17 18 19 20 21 ... 89 90

Commentaires sur ces manuels

Pas de commentaire