Vue de la page 24
Downloading Data from an Instrument 21
8. You can view, print, export, or delete downloaded data by entering the
Data or Last Calibration Windows. These windows are accessed by
clicking Data or Last Calibration along the right side of the Download
Window. See “Viewing, Printing, Exporting, & Deleting Data in the Data
Window” on page 23 or “Last Calibration Button” on page 12.
9. After downloading data from an instrument, you can delete all the data
in the instrument by clicking Clear Logger Data if desired. This will not
delete instrument parameters such as serial number, alarm settings, or
autocalibration settings.
WARNING: If you click Clear Logger Data, all data is erased in the
instrument, but not in your computer’s memory. So it’s
advisable that you download the data from the
instrument first before clearing the data.