Setting the Operational Parameters in Edit Mode 24
2. Press and hold the EDIT ENTER button for 2 seconds. The instrument will
display the Edit Mode screen.
3. Press and release the CAL S button. The instrument will alternate between
the following screens that display the two calibration parameters and their
settings.
4. If you wish to cancel setting the calibration parameters or were just viewing the
parameters to confirm their values, press and release the CAL S button to
return to the home screen.
To continue and change parameters, press and release the EDIT ENTER
button. The Air Sample Time parameter will be displayed. For a description of
the Air Sample Time bump test parameter, see “Bump Test & Calibration
Parameters” on page 16.
B / C / E
AIr
GAS
CAL
45
90
CAL
45
AIR--IN