Rki-instruments SDM-2012 Standalone Configuration Manuel d'utilisateur Page 28

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 112
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 27
23 • Setting the Operational Parameters in Edit Mode
6. To edit a parameter, press and release the EDIT ENTER button when that
parameter is displayed. An “E” will appear to the left of the parameter name.
7. Use the BUMP T and CAL S buttons to set the parameter to the desired
value, then press and release the EDIT ENTER button. The “E” to the right of
parameter name will disappear.
8. Repeat step 6 - step 7 to set any other parameters.
9. When you are done setting the parameters, use the BUMP T button to scroll
past the AUTO--C parameter screen. The screen will indicate that the
parameter changes have been saved.
The instrument will return to the home screen.
Setting the Calibration Parameters
Do the following to set the calibration parameters.
1. Turn on the SDM-2012 with an instrument and establish a connection between
them as described above in "Turning on the SDM-2012 with an Instrument".
The instrument will display the home screen.
30
E AIR--IN
30
30
50
OFF
DECISION
50
12.0
50
25.0
%LEL
%
ppm
ppm
CH4
O2
TRANSMIT
CO
H2S
Vue de la page 27
1 2 ... 23 24 25 26 27 28 29 30 31 32 33 ... 111 112

Commentaires sur ces manuels

Pas de commentaire