Rki-instruments M2A Manuel d'utilisateur Page 24

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 75
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 23
19 M2A Transmitter Operator’s Manual
Oxygen Detector, Partial Pressure, Capillary Type, and CSA Capillary Type
The wiring for all the oxygen detectors is the same. White wire to terminal labeled OXY
+, green wire to terminal labeled OXY -.
Figure 19: Partial Pressure Oxygen Detector Wiring
Figure 20: Capillary Oxygen Detector Wiring
Figure 21: CSA Capillary Type Oxygen Detector Wiring
White
Green
- S +
PWR/SIG
TOXIC OXY
+ - + -
RS 485
A B C
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
FAIL
White
Green
- S +
PWR/SIG
TOXIC OXY
+ - + -
RS 485
A B C
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
FAIL
White
Green
- S +
PWR/SIG
TOXIC OXY
+ - + -
RS 485
A B C
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
FAIL
Vue de la page 23
1 2 ... 19 20 21 22 23 24 25 26 27 28 29 ... 74 75

Commentaires sur ces manuels

Pas de commentaire