71 Replacing or Recharging the Batteries GX-2003 Operator’s Manual
while pushing down the other end to install it into the
compartment.
Figure 6: Installing the Battery Pack
5. Reinstall the battery cover and turn the battery cover release knob
clockwise to lock the cover in place.
Recharging the Battery Pack in the Instrument
CAUTION:To be used only with Ni-MH battery pack p/n 49-1606RK.
Charge only with RKI charger model BC-2003, p/n 49-2160RK,
or RKI charger model BC-2003DC, p/n 49-2161RK. Use of
other rechargeable batteries or chargers or charging of other
rechargeable batteries in the GX-2003 will void the warranty.
1. Plug the power cord of the optional battery charging station into
an AC outlet.
2. Make sure the GX-2003 is off.
3. Place the GX-2003 into the battery charging station as shown in
Figure 7 below so that the metal contacts on the bottom of the unit
come into contact with the metal contacts in the bottom of the
holder in the charging station. When proper contact has been
made, the red LED on the charging station will turn on.