9 M2 Transmitter Operator’s Manual
Figure 7: 65-2495RK H
2
S Detector Non-Explosion Proof
The construction of the various ESM-01 detectors is similar. Each of the ESM-01 detectors
has a different replaceable plug-in sensor that protrudes through the detector housing
cap. Although ESM-01 detectors look the same except for the plug-in detector, the detector
housing which includes the detector socket and conditioning circuitry is specific to a
sensor type. For example, a detector housing from a Cl
2
detector cannot be used with an
NH
3
detector. One of the two detector wires is color coded for each specific detector type.
The ESM-01 detectors include a removable splashguard that screws onto the detector
housing cap. When it is necessary to apply gas to an ESM-01 detector, remove the
splashguard to allow access to the sensor that protrudes through the detector housing cap.
You can then push the appropriate ESM-01 calibration cup onto the sensor.
Figure 8: ESM-01 Type Detector
Cap Gasket
Detector Housing Body
Hydrophobic
Membrane
H2S Plug-in Sensor
Detector
Housing Cap
Detector Housing Body
(different for each detector type)
Cap Gasket
Splash Guard
(removable)
Detector Housing
Cap
Sensor
(different for each detector type)