INSTRUCTION MANUAL
RKI INSTRUMENTS, INC.
MODELS: OX-94, CO-94, HS-94
PORTABLE SINGLE GAS MONITORS
RKI Instruments, Inc.
33248 Central Ave.
Union City, CA 94587
(510) 441-5656
1
Vue de la page 0
1 2 3 4 5 6 ... 12 13

Résumé du contenu

Page 1

INSTRUCTION MANUAL RKI INSTRUMENTS, INC. MODELS: OX-94, CO-94, HS-94 PORTABLE SINGLE GAS MONITORSRKI Instruments, Inc.33248 Centr

Page 2

2. While continuing to hold the AIR Button, press also the MODE button untilinstrument beeps.3. After the beep, release the MODE button but continue t

Page 3

30 or 60 (for H2S) or 150 (for CO). Once you have scrolled to the value of yourcalibration gas, go to step 9.9. Press the MODE switch, and the instru

Page 4

•O2 SensorReplace the O2 sensor when:1. The O2 circuit cannot be set to 0 on an oxygen-free sample.2. The OXY (O2) display does not show 20.9% immedi

Page 5

PRODUCT WARRANTY1/1/2002RKI Instruments, Inc. warranties gas alarm equipment sold by us to be free from defects inmaterials, workmanship, and performa

Page 6

INTRODUCTIONThe RKI Models OX-94, CO-94, and HS-94 (the 94 series) are compact portable gasdetection instruments. They are designed to protect the wor

Page 7

The 94 series instruments are very compact to easily fit in a shirt pocket or on a belt. Italso comes equipped with a lanyard which can be used to han

Page 8

Oxygen SensorThe oxygen sensor is an electrochemical cell, consisting of gold and lead electrodesin an alkaline electrolyte. A fluorocarbon membrane c

Page 9 - CALIBRATION AND MAINTENANCE

CIRCUIT BOARDSThe GX-82 has two circuit boards; the main board, and the sensor/buzzer board. Themain board contains the electronics and the display. T

Page 10

WARNINGIf the Instrument does not respond to these verifications, take it to a known“fresh-air” environment, then follow the zero and calibration proc

Page 11

NoteThe 94 series automatically checks battery voltage during start-up; if the measurementis below 2.1V, the instrument will not operate. During norma

Page 12

4. Auto Zero/SpanTo automatically adjust the zero point for H2S or CO, (or the span to 20.9% foroxygen), take the instrument to a known fresh-air envi

Page 13 - 1/1/2002

SENSOR FAIL This alarm will sound if the sensor is missing orhas an internal open connection. This type ofalarm is indicated by the steady audible ton